MTS6露点仪Moisture Targes Series 6露点仪
MTS6露点仪是一个基于微处理器的,单通道测量气体中的水分含量的露点分析仪。它是可用于设备制造商(OEM)使用,并且适用于要求实时的广泛的工艺条件。MTS6露点仪可以配备GE的M系列露点变送器、TF系列露点变送器。它配备了两个标准报警继电器,一个故障报警继电器,和一个单一的模拟输出。它还具有一个微型SD卡的机载数据记录功能。
MTS6带有LCD显示屏,通过面部前部按键你可以轻松选择需要的露点测量单位,以及其他信息,MTS6露点仪电压从110VAC到240VAC,或者24VDC
Sensor Type:Thin-film aluminum oxide
Moisture Probe Compatibility
Compatible with all GE Sensing M Series aluminum
oxide moisture probes and VeriDri transmitters
Dew/Frost Point Temperature
• Overall range capability: 140°F to –166°F (60°C to –110°C)
• Standard 68°F to –112°F (20°C to –80°C) with data to -166°F (–110°C)
• Extended 140°F to –85°F (60°C to –80°C) with data to 166°F (–110°C)
• VeriDri ranges include:
• –110°C to 20°C
• –110°C to –50°C
• –90°C to 10°C
• –80°C to –30°C
• –80°C to 20°C
• –30°C to 20°C
• –130°F to 70°F
• –130°F to –40°F
• –40°F to 70°F
• –50°F to 50°F
• –100°F to 0°F
Calibrated Accuracy at 77°F (25°C)
• ±3.6°F (±2°C) from –85°F to 50°F (–65°C to 10°C)
• ±5.4°F (±3°C) from –112°F to –86°F (–80°C to –66°C)
Repeatability
• ±0.9°F (±0.5°C) from –85°F to 50°F (–65°C to 10°C)
dew/frost point
• ±1.8°F (±1.0°C) from –112°F to –86°F (–80°C to –66°C)dew/frost point